Toddler Aerosol Study

4 Apr 2007    In vitro analysis of the Singing Arrow Oxyphone Delivery System and VixOne Disposable Nebulizer with Unit Dose Albuterol (2.5mg/3ml)

It was somewhat of a surprise that Oxyphone, with the mouthpiece about 4 cm from the mouth of the model, exceeded the performance of a close-fitting mask .   (Presented at American Thoracic Society meeting in Toronto, May 2008)

 

n = 5

 

OxyphoneÔ

AFM 1

AFM @ 2 2

VMD

4.3 ± 0.1 mm

GSD

2.0

% £ 3mm

37% ± 0.8

% £ 5mm

57% ± 1.2

Duration

6.0

Residual Dose (mg)

1.34 ± 0.13

1.58 ± 0.13

1.32 ± 0.03

Inhaled Dose (mg)

0.337 ± 0.05

0.205 ± 0.01

0.097 ± 0.04

Fine Particle Dose (mg)

0.191 ± 0.03

0.116 ± 0.01

0.055 ± 0.02

Output (FPD/Min)

0.03 ± 0.0

0.02 ± 0.0

0.01 ± 0.0

FPD %

7.6 % ± 1.2

4.6 % ± 0.3

2.2 % ± 0.9

 mean ± sd

1. AFM: Aerosol Face Mask

2. AFM@2: Aerosol Face Mask 2 cm from face

 

 

Powered by AtomicShops